Electrosurgical high frequency ideal for carrying out operations of monopolar cut and coagulated (with different levels of coagulation), bipolar cut and coagulation. During bipolar coagulation there an automatic "Autostart-Autostop" system is activated when coagulation is reached.
You can memorizzae and easily recall a range of 10 different modes and power levels. You can use both electrode plate neutral conductive zone electrodes in bipartite, to control the impedance stability of plate contact with the patient during the surgical operation. The control unit is via the buttons and the front panel display; mains inlet is located on the rear panel. The unit is equipped with an automatic control system to monitor the internal parameters and alert on possible damage or errors. The operating parameters are constantly stored so that the last selected parameters are recalled, every time the unit is switched on or the operative method is changed. The sound level is adjusted, every operator can select its value depending on the working conditions. The units can work through handles with or without pushbuttons but with double foot. You can connect bipolar forceps to the unit for the bipolar functions.
Technical features
Storing job settings
Digital power
Check Control continuously, indicating successful control and storage of errors found
Automatic plate-patient connection
Automatic impedance
Automatic control of excessive time
Automatic level control of power supplied
Bipolar coagulation activation / Silent. Automatic
Consistent with Argon unit
Minimum preselectable power: 0
Increases in unit power from 0W to 50W
step 10 for powers higher than 50W
Digital level display
Selection of power through buttons
Dimensions WxHxD: 470 x 150 x 380 mm
Weight: 18 kg
Selectable mains power (Vac): 115-230
Main Frequency: 50-60 Hz
Fuse 230 Vac (5x20) TIMED: 2xT 4A / 5 °
Fuse 115 Vac (5x20) TIMED: 2xT 8A / 10 °
Electrical power input (VA): 800/1000
Electricity inlet (230 vac) (A): 3.5 / 4.5
Electricity input (115Vac) (A): 7/9
Sound level adjustable, five levels
Integrated self diagnostics
Notice accuracy output power
Electronic plate to patient contact
Permitted use electrodes normal type or bipartite
Storing conditions: last 10
Electrical class (EN 60601-1): ICF
MDD 93 / 42EEC Class: IIb
EN55011 (CISPR 1 liter) Class (class / group): 2 / B
Patient circuit: F
Duty cycle (action / pause) in seconds: 10-30
Output power control by foot-switch or finger
Defibrillation
overheat warning (OVH) 10/2
Grounding bonding
Metallic structure
Lining panels polycarbonate
Complies with EN 60601-1 (1998)
Complying with EN 60601-1-2 (2003)
Complying with EN 60601-2-2 (2000)
Working frequency monopolar: 425 kHz
Maximum output power monopolar pure CUT: 250 W
Impedance: 300 ohms
Maximum power load: 1500 Vpp
Crest factor: 1.6
Maximum output power monopolar coagulation CUT80%: 220 W
Impedance: 300 ohms
Maximum power load: 1500 Vpp
Modulation frequency: 10Khz
Duty cycle: 80%
Crest factor: 1.8
Maximum output power monopolar coagulation CUT60%: 200 W
Impedance: 300 ohms
Maximum power load: 1500 Vpp
Modulation frequency: 10Khz
Duty cycle: 60%
Crest factor: 2.1
Maximum output power monopolar coagulation CUT40%: 150 W
Impedance: 300 ohms
Maximum power load: 1500 Vpp
Modulation frequency: 10Khz
Duty cycle: 40%
Crest factor: 2.6
Maximum power monopolar cut ENHANCED: 150 W
Impedance: 300 ohms
Maximum power load: 1500 Vpp
Modulation frequency: 25.1 KHz
Duty cycle: 5%
Crest factor: 1.6
Maximum output power monopolar coagulation BLEND: 150 W
Impedance: 500 ohms
Maximum power load: 2500 Vpp
Crest factor: 1.6
Maximum power SPEEDY monopolar coagulation: 100 W
Impedance: 500 ohms
Maximum power load: 2500 Vpp
Frequency Modulation: 8-12Khz
Duty cycle: 20%
Do ttore Crest: 3.4
Maximum power monopolar coagulation DEEP: 100 W
Impedance: 500 ohms
Maximum power load: 800 Vpp
Crest factor: 1.6
Maximum power coagulation - spray monopolar SPRAY: 60 W
Impedance: 2000 ohms
Maximum power load: 4500 Vpp
Modulation frequency: 1 kHz
Duty cycle: 20%
Crest factor: 3.4
Maximum power coagulation - spray monopolar SPRAY ARGON: 60 W
Impedance: 2000 ohms
Maximum power load: 4500 Vpp
Modulation frequency: 1 kHz
Crest factor: 3.4
Frequency Bipolar: 525 kHz
Maximum output power bipolar CUT: 120 W
Impedance: 150 ohms
Maximum power load: 800 Vpp
Crest factor: 1.5
Maximum output power bipolar coagulation CUT80%: 100 W
Impedance: 150 ohms
Maximum power load: 800 Vpp
Modulation frequency: 10 Khz
Duty cycle: 80%
Crest factor: 1.7
Maximum output power bipolar coagulation CUT60%: 100 W
Impedance: 150 ohms
Maximum power load: 800 Vpp
Modulation frequency: 10 Khz
Duty cycle: 60%
Crest factor: 2.0
Maximum output power bipolar coagulation CUT40%: 60 W
Impedance: 150 ohms
Maximum power load: 800 Vpp
Modulation frequency: 10 Khz
Duty cycle: 40%
Crest factor: 2.4
Maximum output power bipolar coagulation BLEND: 100 W
Impedance: 300 ohms
Maximum power load: 1100 Vpp
Crest factor: 1..7
Maximum power bipolar coagulation DEEP COAG: 70 W
Impedance: 100 ohms
Maximum power load: 400 Vpp
Crest factor: 1.5
Standard equipment
1 power cord SIEMENS-EEC of 5 m
1 two-button handpiece (autoclavable, cod. 30547)
5 handpieces two buttons (cold sterilizable)
1 Cable for disposable plate
5 disposable plates
2 double waterproof pedal
3 flat electrodes for general surgery courts, 7 cm
3 flat electrodes for general surgery long, 10 cm
3 needle electrodes for general surgery courts, 5.5 cm
3 ball electrodes Ψ 4 mm for general surgery, 6 cm
1 sponge cleaner electrodes
1 user manual (English, Italian)
1 service manual (English, Italian)
Google Traduttore per il Business:Translator ToolkitTraduttore di siti webStrumento a s
|